Microlife Microlife
  • Consumer Products
    • BP B6 Connect_half-cuff Blood Pressure
    • NC200_front Thermometers
    • BreastPump_BC200-front Baby Care
    • NEB NANObasic_half-acc Respiratory Care
    • WS 200 BT_front Weight
    • Software_downloads_allg Software
  • Professional Products
    • WatchBP Office ABI_front_cuff WatchBP Office
    • WatchBP O3 Ambulatory 2G_half_cuff WatchBP O3
    • WatchBP Home A_front_cuff WatchBP Home
    • Cuff-Ankle Cuffs and supplies
    • Software_downloads_WBP Software
  • Clinical Validations
  • Technologies
  • Health Magazine
    • icon_heart_blue Cardiovascular Health
    • icon_Temperature Fever Care
    • icon_airway-treatment Respiratory Care
    • icon_even_heat_blue Flexible Heating
  • Support
    • BP A2 Basic startseite Product Support
    • Software_downloads_allg Home-Use Software
    • Software_downloads_WBP Professional Software
    • Software Developers Developers
  • Company
    • aboutus About us
    • Events News & Events
    • contact Contact
    • History History
ASIA PACIFIC
English
Europe
  • EUROPE - English
  • CZECHIA - Čeština
  • DANMARK - Dansk
  • DEUTSCHLAND - Deutsch
  • ESPAÑA - Español
  • FRANCE - Français
  • HRVATSKA - Hrvatski
  • ITALIA - Italiano
  • LATVIJA - Latviešu
  • LIETUVA - Lietuvių
  • MAGYARORSZÁG - Magyar
  • NEDERLAND - Nederlands
  • ÖSTERREICH - Deutsch
  • POLSKA - Polski
  • PORTUGAL - Português
  • SCHWEIZ - Deutsch
  • SLOVENSKO - Slovenčina
  • SLOVENIJA - Slovenščina
  • SUOMI - Suomi
  • SCHWEIZ - Français
  • UK - English
  • ΕΛΛΆΔΑ - Ελληνικά
  • БЪЛГАРИЯ - Български
  • СРБИЈА - Српски
  • УКРАЇНА - Українська
  • TÜRKIYE - Türkçe
  • РОССИЯ - Русский
  • БЕЛАРУСЬ - Русский
Asia
  • ASIA PACIFIC - English
  • VIỆT NAM - Tiếng Việt
  • ประเทศไทย - ไทย
  • 中国 - 中文
  • 台灣 - 中文
Middle East and North Africa
  • MENA - Arabic عربى
USA, Latin America
  • US - English
  • AMÉRICA DEL SUR - Español
  • COLOMBIA - Español
  • MÉXICO - Español
Menu
Header_WatchBP Validation-Diabetes Flow

Validation of the Microlife BP A3 PC upper arm blood pressure monitor in patients with diabetes mellitus according to the ANSI/AAMI/ISO 81060-2:2013 protocol

Home > Clinical Validations > Validation of the Microlife BP A3 PC upper arm blood pressure monitor in patients with diabetes mellitus according to the ANSI/AAMI/ISO 81060-2:2013 protocol, Beime, 2017
Share

Beate Beimea, Ralf Krügera, Gertrud Hammelb, Peter Bramlagea and Cornelia Deutscha

Objective

The aim of the present study was to validate the blood pressure (BP) measurement device, Microlife BP A3 PC, in patients with diabetes mellitus, according to the ANSI/AAMI/ISO 81060-2:2013 protocol.

Patients and methods

In 85 individuals aged 56–88 years, with predefined criteria for diabetes mellitus, BP measurements on the upper arm were performed alternately using the Microlife BP A3 PC and a standard mercury reference sphygmomanometer. A total of 333 comparisons were included for analysis.

Results

The mean difference between the Microlife BP A3 PC and the reference was −1.5± 6.3 mmHg for systolic BP (SBP) and −1.3± 5.2 mmHg for diastolic BP (DBP) according to criterion 1 of the protocol. For SBP, a total of 209 of the 333 measurements were within the range of 5 mmHg (62.8%), whereas the corresponding numbers for DBP were 232 of 333  (69.7%). For criterion 2, the intraindividual differences for the test device and the reference were −1.50± 4.73 mmHg for SBP and −1.30 ± 4.55 mmHg for DBP, thus being within the defined ranges provided by the protocol.

Conclusion

The Microlife BP A3 PC fulfilled the requirements of criteria 1 and 2 of the ANSI/AAMI/ISO 81060-2:2013 protocol and can also be recommended for BP measurement in diabetic patients.

Click here to download the abstract in pdf format.